產(chǎn)品簡介
product
產(chǎn)品分類These enhanced performance sensors are manufactured to exacting tolerances using high quality, rugged materials for demanding applications including ultra-pure water, clean-in-place (CIP), and boiler/condensate monitoring. Each sensor is tested to determine its unique, absolute four-digit cell constant. Simply key in this constant (Hach’s easy DRY-CAL™ method) when configuring the analyzer to ensure the highest possible measuring accuracy. Also, each sensor has a Pt 1000 RTD temperature element built into its tip for exceptionally fast response to changes in temperature with ±0.1ºC accuracy.
These enhanced performance sensors measure from theoretically pure water (0.057 µS/cm or 18.2 MΩ) up to 200,000 µS/cm. Hach’s sc Digital Controllers accept multiple digital sensor inputs, and can be user-set to measure conductivity, resistivity,TDS, salinity, or one of six calculated measurements.
Compression fitting sensors—Feature titanium electrodes and a compression fitting for universal installation with up to 4 inches (102 mm) insertion depth. The 1/2-inch or 3/4-inch male NPT compression fittings are offered in Kynar® (PVDF) or 316 stainless steel. A longer version of this sensor is available for use with a 316 stainless steel ball valve hardware assembly to insert/retract the sensor from the process without stopping the flow. The longer version can also be used for insertion through a compression fitting. Maximum insertion depth is 7 inches (178 mm).
There are no complicated wiring or set up procedures with any Hach sc controller. Just plug in any combination of Hach digital sensors and it's ready to use—it’s “plug and play.”
Accuracy:± 2 % of reading above 200 μS/cm
Cable Length:6 m fixed cable (6 wire, 4 conductors and two isolated shield wires)
Diameter:0.5 in ( 12.7 mm )
Flow:0 to 3 m/s maximum, fully immersed
Length:7.6 in ( 193.5 mm )
Material:Compression Fitting: KYNAR (PVDF)
Measurement Range 2:0.002 to 20 MΩ resistivity
Operating Temperature Range:-20 to 200 °C
Pressure Limit:1.7 bar at 150 °C
Range:0 to 100 μS/cm conductivity
Repeatability:± 0.5 % of reading
Response Time T90:30 s
Sensor Immersion Depth:155 mm
Sensor Thread:NPT
Temperature Sensor:Pt1000 RTD
Transmission Distance:max. 100 m